Location

2008/04/25 17:31:00 38.450 -87.870 13.0 4.2 Illinois

Arrival Times (from USGS)

Arrival time list

Felt Map

USGS Felt map for this earthquake

USGS Felt reports page for

Focal Mechanism

 SLU Moment Tensor Solution
 2008/04/25 17:31:00 38.450 -87.870 13.0 4.2 Illinois
 
 Best Fitting Double Couple
    Mo = 4.79e+21 dyne-cm
    Mw = 3.72 
    Z  = 13 km
     Plane   Strike  Dip  Rake
      NP1      204    85   170
      NP2      295    80     5
 Principal Axes:
   Axis    Value   Plunge  Azimuth
     T   4.79e+21     11     159
     N   0.00e+00     79     358
     P  -4.79e+21      4     250



 Moment Tensor: (dyne-cm)
    Component  Value
       Mxx     3.48e+21
       Mxy    -3.07e+21
       Mxz    -7.05e+20
       Myy    -3.62e+21
       Myz     5.85e+20
       Mzz     1.43e+20
                                                     
                                                     
                                                     
                                                     
                     ##############                  
                 ###################---              
              ####################--------           
             ####################----------          
           #####################-------------        
          #####################---------------       
         ---###################----------------      
        -----------###########------------------     
        ----------------#####-------------------     
       ------------------------------------------    
       --------------------#####-----------------    
       -------------------#########--------------    
       ------------------##############----------    
           --------------#################------     
         P -------------####################----     
           ------------#######################-      
          ------------########################       
           -----------#######################        
             --------######################          
              ------#############   ######           
                 ---############# T ###              
                     ############                    
                                                     
                                                     
                                                     

 Harvard Convention
 Moment Tensor:
      R          T          F
  1.43e+20  -7.05e+20  -5.85e+20 
 -7.05e+20   3.48e+21   3.07e+21 
 -5.85e+20   3.07e+21  -3.62e+21 


Details of the solution is found at

http://www.eas.slu.edu/eqc/eqc_mt/MECH.NA/20080425173100/index.html
        

Preferred Solution

The preferred solution from an analysis of the surface-wave spectral amplitude radiation pattern, waveform inversion and first motion observations is

      STK = 295
      DIP = 80
     RAKE = 5
       MW = 3.72
       HS = 13.0

The waveform inversion is preferred.

Moment Tensor Comparison

The following compares this source inversion to others
SLU
 SLU Moment Tensor Solution
 2008/04/25 17:31:00 38.450 -87.870 13.0 4.2 Illinois
 
 Best Fitting Double Couple
    Mo = 4.79e+21 dyne-cm
    Mw = 3.72 
    Z  = 13 km
     Plane   Strike  Dip  Rake
      NP1      204    85   170
      NP2      295    80     5
 Principal Axes:
   Axis    Value   Plunge  Azimuth
     T   4.79e+21     11     159
     N   0.00e+00     79     358
     P  -4.79e+21      4     250



 Moment Tensor: (dyne-cm)
    Component  Value
       Mxx     3.48e+21
       Mxy    -3.07e+21
       Mxz    -7.05e+20
       Myy    -3.62e+21
       Myz     5.85e+20
       Mzz     1.43e+20
                                                     
                                                     
                                                     
                                                     
                     ##############                  
                 ###################---              
              ####################--------           
             ####################----------          
           #####################-------------        
          #####################---------------       
         ---###################----------------      
        -----------###########------------------     
        ----------------#####-------------------     
       ------------------------------------------    
       --------------------#####-----------------    
       -------------------#########--------------    
       ------------------##############----------    
           --------------#################------     
         P -------------####################----     
           ------------#######################-      
          ------------########################       
           -----------#######################        
             --------######################          
              ------#############   ######           
                 ---############# T ###              
                     ############                    
                                                     
                                                     
                                                     

 Harvard Convention
 Moment Tensor:
      R          T          F
  1.43e+20  -7.05e+20  -5.85e+20 
 -7.05e+20   3.48e+21   3.07e+21 
 -5.85e+20   3.07e+21  -3.62e+21 


Details of the solution is found at

http://www.eas.slu.edu/eqc/eqc_mt/MECH.NA/20080425173100/index.html
	

Waveform Inversion

The focal mechanism was determined using broadband seismic waveforms. The location of the event and the and stations used for the waveform inversion are shown in the next figure.
Location of broadband stations used for waveform inversion

The program wvfgrd96 was used with good traces observed at short distance to determine the focal mechanism, depth and seismic moment. This technique requires a high quality signal and well determined velocity model for the Green functions. To the extent that these are the quality data, this type of mechanism should be preferred over the radiation pattern technique which requires the separate step of defining the pressure and tension quadrants and the correct strike.

The observed and predicted traces are filtered using the following gsac commands:

hp c 0.02 n 3
lp c 0.10 n 3
The results of this grid search from 0.5 to 19 km depth are as follow:

           DEPTH  STK   DIP  RAKE   MW    FIT
WVFGRD96    0.5   295    75    20   3.37 0.2494
WVFGRD96    1.0   295    75    20   3.42 0.2733
WVFGRD96    2.0   295    80    15   3.48 0.3128
WVFGRD96    3.0   295    90    20   3.53 0.3325
WVFGRD96    4.0   295    90    20   3.56 0.3517
WVFGRD96    5.0   115    85   -20   3.58 0.3721
WVFGRD96    6.0   115    85   -20   3.60 0.3875
WVFGRD96    7.0   115    85   -20   3.61 0.3988
WVFGRD96    8.0   115    90   -15   3.63 0.4103
WVFGRD96    9.0   115    90   -15   3.65 0.4224
WVFGRD96   10.0   295    90    15   3.67 0.4352
WVFGRD96   11.0   295    85    10   3.69 0.4493
WVFGRD96   12.0   295    85    10   3.70 0.4583
WVFGRD96   13.0   295    80     5   3.72 0.4625
WVFGRD96   14.0   295    80     5   3.73 0.4624
WVFGRD96   15.0   295    80     5   3.74 0.4592
WVFGRD96   16.0   295    80     5   3.74 0.4520
WVFGRD96   17.0   295    80     5   3.75 0.4408
WVFGRD96   18.0   295    80     5   3.75 0.4262
WVFGRD96   19.0   295    80     5   3.76 0.4102
WVFGRD96   20.0   295    80     5   3.76 0.3931
WVFGRD96   21.0   115    90   -10   3.76 0.3783
WVFGRD96   22.0   295    90    10   3.76 0.3641
WVFGRD96   23.0   295    90    10   3.76 0.3498
WVFGRD96   24.0   295    90    10   3.76 0.3358
WVFGRD96   25.0   295    90    10   3.76 0.3226
WVFGRD96   26.0   295    90    10   3.75 0.3094
WVFGRD96   27.0   115    85   -10   3.75 0.2987
WVFGRD96   28.0   295    90    10   3.75 0.2877
WVFGRD96   29.0   295    90    10   3.75 0.2779

The best solution is

WVFGRD96   13.0   295    80     5   3.72 0.4625

The mechanism correspond to the best fit is
Figure 1. Waveform inversion focal mechanism

The best fit as a function of depth is given in the following figure:

Figure 2. Depth sensitivity for waveform mechanism

The comparison of the observed and predicted waveforms is given in the next figure. The red traces are the observed and the blue are the predicted. Each observed-predicted componnet is plotted to the same scale and peak amplitudes are indicated by the numbers to the left of each trace. The number in black at the rightr of each predicted traces it the time shift required for maximum correlation between the observed and predicted traces. This time shift is required because the synthetics are not computed at exactly the same distance as the observed and because the velocity model used in the predictions may not be perfect. A positive time shift indicates that the prediction is too fast and should be delayed to match the observed trace (shift to the right in this figure). A negative value indicates that the prediction is too slow. The bandpass filter used in the processing and for the display was

hp c 0.02 n 3
lp c 0.10 n 3
Figure 3. Waveform comparison for depth of 8 km
Focal mechanism sensitivity at the preferred depth. The red color indicates a very good fit to thewavefroms. Each solution is plotted as a vector at a given value of strike and dip with the angle of the vector representing the rake angle, measured, with respect to the upward vertical (N) in the figure.

Surface-Wave Focal Mechanism

The following figure shows the stations used in the grid search for the best focal mechanism to fit the surface-wave spectral amplitudes of the Love and Rayleigh waves.
Location of broadband stations used to obtain focal mechanism from surface-wave spectral amplitudes

The surface-wave determined focal mechanism is shown here.


  NODAL PLANES 

  
  STK=      25.00
  DIP=      90.00
 RAKE=    -170.00
  
             OR
  
  STK=     295.00
  DIP=      80.00
 RAKE=      -0.01
 
 
DEPTH = 13.0 km
 
Mw = 3.84
Best Fit 0.8002 - P-T axis plot gives solutions with FIT greater than FIT90

First motion data

The P-wave first motion data for focal mechanism studies are as follow:

Sta Az(deg)    Dist(km)   First motion

Surface-wave analysis

Surface wave analysis was performed using codes from Computer Programs in Seismology, specifically the multiple filter analysis program do_mft and the surface-wave radiation pattern search program srfgrd96.

Data preparation

Digital data were collected, instrument response removed and traces converted to Z, R an T components. Multiple filter analysis was applied to the Z and T traces to obtain the Rayleigh- and Love-wave spectral amplitudes, respectively. These were input to the search program which examined all depths between 1 and 25 km and all possible mechanisms.
Best mechanism fit as a function of depth. The preferred depth is given above. Lower hemisphere projection

Pressure-tension axis trends. Since the surface-wave spectra search does not distinguish between P and T axes and since there is a 180 ambiguity in strike, all possible P and T axes are plotted. First motion data and waveforms will be used to select the preferred mechanism. The purpose of this plot is to provide an idea of the possible range of solutions. The P and T-axes for all mechanisms with goodness of fit greater than 0.9 FITMAX (above) are plotted here.


Focal mechanism sensitivity at the preferred depth. The red color indicates a very good fit to the Love and Rayleigh wave radiation patterns. Each solution is plotted as a vector at a given value of strike and dip with the angle of the vector representing the rake angle, measured, with respect to the upward vertical (N) in the figure. Because of the symmetry of the spectral amplitude rediation patterns, only strikes from 0-180 degrees are sampled.

Love-wave radiation patterns

Rayleigh-wave radiation patterns

Broadband station distribution

The distribution of broadband stations with azimuth and distance is

Sta Az(deg)    Dist(km)   
WCI	  100	  140
BLO	   55	  142
SIUC	  236	  144
SLM	  276	  207
UTMT	  201	  250
WVT	  179	  258
PVMO	  216	  278
PBMO	  231	  292
CCM	  263	  299
PLAL	  183	  385
MPH	  207	  413
TZTN	  118	  437
OXF	  198	  458
AAM	   39	  557
SCIA	  312	  595
LRAL	  172	  606
BLA	  100	  670
MIAR	  231	  670
GOGA	  143	  686
MCWV	   77	  707
GLMI	   20	  758
KSU1	  278	  763
ERPA	   56	  784
BRAL	  174	  811
COWI	  353	  856
NHSC	  128	  913
CBN	   88	  918
ECSD	  311	  939
SDMD	   80	  962
MVL	   77	 1009
CBKS	  276	 1033
BINY	   64	 1091
EYMN	  346	 1095
HKT	  220	 1193
PAL	   72	 1229
FRNY	   55	 1382
JCT	  235	 1406
RSSD	  300	 1490
PHWY	  287	 1535
ISCO	  281	 1539
SDCO	  273	 1546
SMCO	  279	 1658
RWWY	  288	 1685
ANMO	  262	 1703
DGMT	  316	 1721
MVCO	  272	 1817
RLMT	  300	 1920

Waveform comparison for this mechanism

Since the analysis of the surface-wave radiation patterns uses only spectral amplitudes and because the surfave-wave radiation patterns have a 180 degree symmetry, each surface-wave solution consists of four possible focal mechanisms corresponding to the interchange of the P- and T-axes and a roation of the mechanism by 180 degrees. To select one mechanism, P-wave first motion can be used. This was not possible in this case because all the P-wave first motions were emergent ( a feature of the P-wave wave takeoff angle, the station location and the mechanism). The other way to select among the mechanisms is to compute forward synthetics and compare the observed and predicted waveforms.

The fits to the waveforms with the given mechanism are show below:

This figure shows the fit to the three components of motion (Z - vertical, R-radial and T - transverse). For each station and component, the observed traces is shown in red and the model predicted trace in blue. The traces represent filtered ground velocity in units of meters/sec (the peak value is printed adjacent to each trace; each pair of traces to plotted to the same scale to emphasize the difference in levels). Both synthetic and observed traces have been filtered using the SAC commands:

hp c 0.02 n 3
lp c 0.10 n 3

Discussion

The Future

Should the national backbone of the USGS Advanced National Seismic System (ANSS) be implemented with an interstation separation of 300 km, it is very likely that an earthquake such as this would have been recorded at distances on the order of 100-200 km. This means that the closest station would have information on source depth and mechanism that was lacking here.

Acknowledgements

Dr. Harley Benz, USGS, provided the USGS USNSN digital data. The digital data used in this study were provided by Natural Resources Canada through their AUTODRM site http://www.seismo.nrcan.gc.ca/nwfa/autodrm/autodrm_req_e.php, and IRIS using their BUD interface.

Thanks also to the many seismic network operators whose dedication make this effort possible: University of Alaska, University of Washington, Oregon State University, University of Utah, Montana Bureas of Mines, UC Berkely, Caltech, UC San Diego, Saint L ouis University, Universityof Memphis, Lamont Doehrty Earth Observatory, Boston College, the Iris stations and the Transportable Array of EarthScope.

Appendix A


Spectra fit plots to each station

Velocity Model

The CUS used for the waveform synthetic seismograms and for the surface wave eigenfunctions and dispersion is as follows:

MODEL.01
CUS Model with Q from simple gamma values
ISOTROPIC
KGS
FLAT EARTH
1-D
CONSTANT VELOCITY
LINE08
LINE09
LINE10
LINE11
  H(KM) VP(KM/S) VS(KM/S) RHO(GM/CC)   QP   QS  ETAP  ETAS  FREFP  FREFS
  1.0000  5.0000  2.8900  2.5000 0.172E-02 0.387E-02 0.00  0.00  1.00  1.00 
  9.0000  6.1000  3.5200  2.7300 0.160E-02 0.363E-02 0.00  0.00  1.00  1.00 
 10.0000  6.4000  3.7000  2.8200 0.149E-02 0.336E-02 0.00  0.00  1.00  1.00 
 20.0000  6.7000  3.8700  2.9020 0.000E-04 0.000E-04 0.00  0.00  1.00  1.00 
  0.0000  8.1500  4.7000  3.3640 0.194E-02 0.431E-02 0.00  0.00  1.00  1.00 

Quality Control

Here we tabulate the reasons for not using certain digital data sets

The following stations did not have a valid response files:

DATE=Fri Apr 25 17:04:53 CDT 2008

Last Changed 2008/04/25