|
|
Multiflow Automated Water Analyses Topics discussed below:
1. I Multiflow set-up: 1. Open the red-handled valve to the pressurized line behind
the Multiflow. (Assuming that the green valve is already opened on
the line and the air pressure is set to 70 psi).
2. Make sure the power plug and computer plug are properly plugged into the back of the Multiflow unit. Turn on the black switch that is located just above the power plug (by depressing the top half of switch).
3. Make sure He cylinder
in other room is open and that the pressure on the gauge farthest from the
cylinder is set at 400 kPa. Turn on the furnaces for the 4-way and 6-way
valves (VW and VV valves, respectively) and the GC column in the Multiflow
unit. 4. Turn up the gas flow through the Reference gas side of the
TCD (at the end of the GC Column) so that 15 ml/min of He is flowing thru
the vent on top of the Gilson unit. Need to have the VU valve set
so that the green box on the computer screen is down (closed). Check
the flow rate by putting the bottom of the Gilman flow meter gently into
the black cylinder. Be gentle with the end of the flow meter when you
attach and detach it to/from the black cylinder. The end comes off
easily. If it comes off, the ball will fall out.
5. Turn up the gas flow through the GC Column so that 15 ml/min
of He is flowing thru the tent on top of the Gilson unit. Need to
have the VU valve set so that the green box on the computer screen is up
(open). At the time of installation of the machine, the value on the
gauge was 19.1; this valuee will vary slightly.
6. Turn the black-handled valve behind the Gilson unit so that
it is parallel to thesteel line (pointing toward the red-handled valve).
This will let in the CO2-He mixture from the tank in the other room.
(Make sure the black-handled valve in the other room is turned toward the
CO2-He tank and not the H2-He tank.) Turn up the gas flow through the sample
vessel so that 40? ml/min of CO2-He is flowing thru the needle. Check
the flow rate by
7. The flush of the needle .......
8. Make sure the needle assembly
is set up for water analyses.
9. Make sure the needle will properly locate itself relative to the sample tray by initializing the Gilson unit (i.e., the unit that controls the needle). You must check this visually by doing the following commands on the computer.... (INLET => INITIALIZE GILSON (moves needle up and down) => PROGRAM => RUN => type in 1,1,1 into the Parameter line => double click on GMOVE. The needle will move the 1st rack and position itself over X=1, Y=1 sample position. If you forgot to input the parameters, you will get an error message for lines 55, 56, and 57 in the GMOVE.seq file. Move the needle down close to the top of the sample vessel by using the GDOWN command. First go to ANALYSIS => PARAMETER FILE EDIT => HEATED SAMPLE TRAY, 1/10 mm => EDIT => NEEDLE DOWN POSITION => input 760 (which will move the needle down 1.5 cm from its "up" or rest position) => OK. Then go to.... PROGRAM => RUN => GDOWN ... and the needle will move down. If the needle appears to be centered on the sample hole, then change the needle down position to 860 and then do another GDOWN command.... Repeat the process of changing the needle down position until you reach 1110. Then MAKE SURE to lift the needle back up by going to ... PROGRAM => RUN => GUP. 10. Turn on the heater to the sample tray by plugging in the
plug on the back of the tray. The temperature of the tray should be
set to 40 C. To check this, push the green button on the left side of the
Sample Temperature gauge and the temperature set point will be displayed.
This is the temperature to which the heat will obtain when it reaches equilibrium.
If the set point is not 40 C, then hold down the green button on the left
and then scroll up or down with the two green buttons on the right.
1. For OXYGEN analysis, pipette 200 microliters of
water into a reaction vessel and then immediately screw on the lid in order
to minimize the loss of water by evaporation. Please be careful with the pipette
-- such pipettes are easy to break. Gently put the tip on, but be careful
to just hold on to the broader end of the tip rather than at the point (in
order to avoid any contamination of the sample). To get the tip off
the pipetter, push gently the green bar down toward the tip. This
will push off the tip. DO NOT PULL THE TIP OFF WITH YOUR HAND.
For HYDROGEN analysis, pipette 200
microliters of water into vessel, add a coil of Hokko beads with tweezers,
and then immediately screw on a lid. DO NOT touch the Hokko coil or
beads with your fingers in order to not contaminate them! Please be
careful with the pipette -- such pipettes are easy to break. Gently
put the tip on, but be careful to just hold on to the broader end of the
tip rather than at the point (in order to avoid any contamination of the
sample). To get the tip off the pipetter, push gently the green bar
down toward the tip. This will push off the tip. DO NOT PULL
THE TIP OFF WITH YOUR HAND.
For both oxygen and hydrogen analyses, you need to analyze standard
waters in addition to your samples. The primary inhouse standard is
known as SLU TAP H2O. There is 10 liter bottle of this standard
and numerous smaller aliquots of this standard in glass bottles. Use
these glass bottles for day-to-day use. There are several other inhouse
water standards that you can use periodically. The isotopic values
of these inhouse standards have been calibrated to VSMOW and GISP standards.
2. Load your samples in the metal sample tray on the Multiflow unit.
3. Place the metal cover over the top of the samples and
gently lower down until it rests again the metal sample holder. You
might have to gently rock the cover in order to get all the sample caps aligned
with the holes in the cover. Then remove the black-handled screws, and
then put on the foam and metal covers (though the covers are not necessary).
The covers are designed to insure uniform temperature among the samples.
CO2Oxygen isotope analyses: 1. On the computer, go to ANALYSIS => PARAMETER FILE
EDIT => MULTIFLOW PARAMETER FILE => EDIT PARAMETER window. Input
your desired parameters and then save the settings. For example, .....
Then hit OK to exit the parameter file. 2. Go to ANALYSIS => METHOD SETUP => LOAD => CO2MULT (in comment line). Go to SEQUENCE and make sure the START SEQUENCE is CSTART and the END SEQUENCE is GILEND. Then click OK => OK (SAVE). 3. Go to... ANALYSIS => EDIT AUTORUN => NEW =>
input a filename that contains 8 characters or less => OK =>
Rack Type to Heated_R check mark AutoRea then for each sample Then CLOSE the AutoRun Batch window.
DO NOT Run the analysis at this time unless you know that all the gases
are flowing properly through the Multiflow and Isoprime.
5. Turn the black Whitey valve on the stainless steel line
behind the Multiflow unit to CO2-He mixture.
6. Can check to make sure that enough CO2-He gas (for flushing the vessel and equilibrating the CO2 gas with the water) will flow through the needle by attaching the soap-filled glass and plastic tubing to the end of the needle. Insert the needle through the septum in the plastic tubing, put some Snoop soap into the glass tubing, compress the bottom of the plastic tubing to spread the soap across the tube, and the monitor the flow rate of the CO2-He mixture. The "VT" valve on the computer monitor must be open for the CO2-He mixture to flow through the needle. The black-knobbed valve on the back of the Multiflow controls the flow rate of the CO2-He mixture through the needle. The flow rate should be 40 ml/minute.
7. Can check to make sure that enough He gas (for pressurizing
the vessel after equlibration) will flow through the needle by attaching
the soap-filled glass and plastic tubing to the end of the needle.
Insert the needle through the septum in the plastic tubing, put some Snoop
soap into the glass tubing, compress the bottom of the plastic tubing to
spread the soap across the tube, and the monitor the flow rate. The
"VB" valve on the computer monitor must be open for the He to flow through
the needle. The "Sample Vessel" valve on the front of the Multiflow
controls the flow rate of the He through the needle. Set this to be
2.9 psi.
8. Go to ... ANALYSIS => EDIT AUTORUN => PREPARED => locate your batch file and double click => RUN. The computer will then take over and do the analyses. You can CLOSE the Auton Batch window. 1. On the computer, go to ANALYSIS => PARAMETER FILE EDIT
=> MULTIFLOW PARAMETER FILE => EDIT PARAMETER window. Input your
desired parameters and then save the settings. For example, .....
2. Go to ANALYSIS => METHOD SETUP => LOAD => H2MULT (in comment line) => OK (SAVE) 3. If you have more than one sample to analyze, then go to...
ANALYSIS => EDIT AUTORUN => NEW => input a filename that contains
8 characters or less => OK =>
Rack Type to Heated_R check mark AutoRea then for each sample Then CLOSE the AutoRun Batch window.
DO NOT Run the analysis at this time unless you know that all the gases
are flowing properly through the Multiflow and Isoprime. 5. Check the TCD is set at 100 and a zero of 0.5 6. Turn the valve on stainless steel line behind the Multiflow
unit to H2-He mixture.
7. Can check to make sure that enough H2-He gas (for flushing the vessel and equilibrating the H2 gas with the water) will flow through the needle by attaching the soap-filled glass and plastic tubing to the end of the needle. Insert the needle through the septum in the plastic tubing, put some Snoop soap into the glass tubing, compress the bottom of the plastic tubing to spread the soap across the tube, and the monitor the flow rate of the H2-He mixture. The "VT" valve on the computer monitor must be open for the H2-He mixture to flow through the needle. The black-knobbed valve on the back of the Multiflow controls the flow rate of the H2-He mixture through the needle. The flow rate should be 40 ml/minute.
8. Can check to make sure that enough He gas (for pressurizing
the vessel after equlibration) will flow through the needle by attaching
the soap-filled glass and plastic tubing to the end of the needle.
Insert the needle through the septum in the plastic tubing, put some Snoop
soap into the glass tubing, compress the bottom of the plastic tubing to
spread the soap across the tube, and the monitor the flow rate. The
"VB" valve on the computer monitor must be open for the He to flow through
the needle. The "Sample Vessel" valve on the front of the Multiflow
controls the flow rate of the He through the needle. Set this to be
2.9 psi.
9. Make sure the Isoprime is set up to do hydrogen analyses. Do the following steps. "Center H" scan, an H3 calibration, and an H2 stability. 10. Go to ... ANALYSIS => EDIT AUTORUN => PREPARED => locate your batch file and double click => RUN. The computer will then take over and do the analyses. Dissolved inorganic carbon analyses: Multiflow Sample Vessel Pressure Acid Delivery Dic. Page 125 (multiflow) (add drawing here) * A = sample He Using Different Racks Calibrating reference gases after changing the tanks: 1. A new tank of CO2 Reference Gas: Go to ... ANALYSIS => METHOD SETUP => CO2MULT => SPECIES => choose CO2=> EDIT. Put in zeros for the values of carbon and oxygen. Then run a series of carbonate standards with a wide range of isotopic values. Plot the results such that the measured values are on the X-axis and the accepted isotopic values are on the Y-axis. The slope of the regression line should be 1.000. The Y-intercept is the isotopic value of the CO2 reference gas in the tank. This value needs to be inputted into the computer program. Go back to ... ANALYSIS => METHOD SETUP => CO2MULT => SPECIES => choose CO2=> EDIT => and input the isotopic values into the boxes. These values are relative to next to "With Respect To" choose PDB => OK. 2. A new tank of H2 Reference Gas: The OS2 software does
not handle hydrogen analyses in the same way it does the CO2. Consequently,
you must use the H2Utils software for putting inthe reference gas isotopic
values etc. Startup the h2utils program by double clicking the icon
on the computer screen. Go to ... FILE => PARAMETERS => REFERENCE
DETAILS => LABORAROTY REFERENCE GAS box and input zero into the Enrichment
(per mil) box and make it with respect to SMOW. Click OK and then run
a series of water standards for hydrogen. Plot the results such that
the measured values are on the X-axis and the accepted isotopic values are
on the Y-axis. The slope of the regression line should be 1.000.
The Y-intercept is the isotopic value of the H2 reference gas in the tank.
Go back to h2utils program FILE => PARAMETERS => REFERENCE DETAILS =>
LABORAROTY REFERENCE GAS box and input the value into the Enrichment (per
mil) box and make it with respect to SMOW. Also input the values you
obtained for VSMOW and SLAP. The program will then give you the "stretch
factor". This value will be printed out with the results of each future analysis.
Click OK and the results provided by the computer should now be corrected
properly for the reference gas. Updated August, 2001 |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| ile => |